Oblique toes X-ray positioning

An extra foot view, used when a routine one is not enough. The patient lies on their back, the beam passes at an oblique angle, straight on to the receptor, and the centring point is the metatarsophalangeal joint of the toe.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Foot

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. From the AP, raise the medial side 30–45°.

Centring & collimation

Central ray

oblique

Centre: the metatarsophalangeal joint of the toe.

Beam: oblique, perpendicular to the receptor.

Collimate: the toes

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmlandscape
PlacementTabletopNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS460.50.5
S460.50.5
M460.630.5
L4610.5

These start from a classic film-screen chart (46 kV, 10 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Phalanges separated.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
OverlapToo little rotationRotate further.

Common questions

Where do you centre for a toes oblique X-ray?
The centring point is the metatarsophalangeal joint of the toe.
What distance and beam angle does a toes oblique need?
100 cm from the tube to the receptor. The beam passes at an oblique angle, straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, landscape, on the tabletop. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a toes oblique?
Overlap. Usual cause: too little rotation. The fix: rotate further.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a toes oblique in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.