AP toes X-ray positioning

An extra foot view, used when a routine one is not enough. The patient lies on their back, the beam passes front to back (AP), tilted 10° towards the head, and the centring point is the metatarsophalangeal joint of the toe.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Foot

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Sole flat on the detector, knee bent.

Centring & collimation

Central ray

AP

Centre: the metatarsophalangeal joint of the toe.

Beam: AP, angled 10° toward-head.

Collimate: the toes and metatarsal heads

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmlandscape
PlacementTabletopNo grid
Angulation10°toward-head

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS460.50.5
S460.50.5
M460.630.5
L4610.5

These start from a classic film-screen chart (46 kV, 10 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Interphalangeal joints open.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Joints closedNo angleAngle toward the heel.

Common questions

Where do you centre for a toes AP X-ray?
The centring point is the metatarsophalangeal joint of the toe.
What distance and beam angle does a toes AP need?
100 cm from the tube to the receptor. The beam passes front to back (AP), tilted 10° towards the head.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, landscape, on the tabletop. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a toes AP?
Joints closed. Usual cause: no angle. The fix: angle toward the heel.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a toes ap in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.