Axial calcaneus X-ray positioning (prone)

An extra foot view, used when a routine one is not enough. The patient lies face down, the beam runs along the length of the part (axial), tilted 60° towards the feet, and the centring point is the heel.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Foot

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Prone. Prone with the toes against the bucky.

Centring & collimation

Central ray

axial

Centre: the heel.

Beam: axial, angled 60° toward-feet.

Collimate: the calcaneus

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmportrait
PlacementErect (wall) buckyGrid in
Angulation60°toward-feet

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS7020.8
S703.21.25
M7041.6
L706.32.5

These start from a classic film-screen chart (70 kV, 32 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The calcaneus is shown axially.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
ForeshortenedAngle wrongCheck 60°.

Common questions

Where do you centre for a heel semiaxial (prone) X-ray?
The centring point is the heel.
What distance and beam angle does a heel semiaxial (prone) need?
100 cm from the tube to the receptor. The beam runs along the length of the part (axial), tilted 60° towards the feet.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, portrait, in the wall bucky. A grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a heel semiaxial (prone)?
Foreshortened. Usual cause: angle wrong. The fix: check 60°.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a heel semiaxial (prone) in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.