Lateral toes X-ray positioning

An extra foot view, used when a routine one is not enough. The patient lies on their side, the beam passes from side to side, straight on to the receptor, and the centring point is the proximal interphalangeal joint of the toe.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Foot

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Lateral recumbent. The injured toe isolated with tape, the others pulled away.

Centring & collimation

Central ray

lateral

Centre: the proximal interphalangeal joint of the toe.

Beam: lateral, perpendicular to the receptor.

Collimate: the toe

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmlandscape
PlacementTabletopNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS460.50.5
S460.50.5
M460.630.5
L4610.5

These start from a classic film-screen chart (46 kV, 10 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The toe is clear of the others.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Toes overlapNot separatedTape the others away.

Common questions

Where do you centre for a toes lateral X-ray?
The centring point is the proximal interphalangeal joint of the toe.
What distance and beam angle does a toes lateral need?
100 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, landscape, on the tabletop. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a toes lateral?
Toes overlap. Usual cause: not separated. The fix: tape the others away.