Lateral pelvis X-ray positioning

An extra pelvis view, used when a routine one is not enough. The patient lies on their side, the beam passes from side to side, straight on to the receptor, and the centring point is the greater trochanter.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Pelvis

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Lateral recumbent. Lateral recumbent with the hips and knees flexed together.

Centring & collimation

Central ray

lateral

Centre: the greater trochanter.

Beam: lateral, perpendicular to the receptor.

Collimate: the whole pelvis in lateral

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor35 × 43 cmportrait
PlacementTable buckyGrid in
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: Suspend respiration.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS903212.5
S905020
M906325
L9010040

These start from a classic film-screen chart (90 kV, 100 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Acetabula superimposed.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
RotationHips not stackedStack the hips and knees.

Common questions

Where do you centre for a pelvis lateral X-ray?
The centring point is the greater trochanter.
What distance and beam angle does a pelvis lateral need?
100 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor.
What size receptor, and do you need a grid?
Use a 35 × 43 cm receptor, portrait, in the table bucky. A grid is needed.
What should the patient do with their breathing?
Ask them to stop breathing for the moment of the exposure.
What goes wrong most often on a pelvis lateral?
Rotation. Usual cause: hips not stacked. The fix: stack the hips and knees.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a pelvis lateral in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.