AP urinary bladder and inner pelvis X-ray positioning
One of the routine pelvis views. The patient lies on their back, the beam passes front to back (AP), tilted 20° towards the feet, and the centring point is 4 cm above the symphysis.
By the ScanFlo Editorial Team · Updated
In this section · X-ray positioning
- X-ray positioning guides
- Skull positioning
- Face & sinuses positioning
- Soft tissue neck positioning
- C-spine positioning
- Chest positioning
- Shoulder positioning
- Humerus positioning
- Elbow positioning
- Forearm positioning
- Wrist positioning
- Hand positioning
- T-spine positioning
- Abdomen positioning
- L-spine positioning
- Sacrum & SI joints positioning
- Pelvis positioning
- Hip positioning
- Femur positioning
- Knee positioning
- Tibia & fibula positioning
- Ankle positioning
- Foot positioning
Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.
Patient position
Supine. Lay the patient supine as for the abdomen.
Centring & collimation
Central ray
APCentre: 4 cm above the symphysis.
Beam: AP, angled 20° caudad.
Collimate: the bladder and pelvic inlet
Distance, receptor & grid
| Setting | Value | Notes |
|---|---|---|
| SID | 100 cm | |
| Receptor | 24 × 30 cm | landscape |
| Placement | Table bucky | Grid in |
| Angulation | 20° | caudad |
Marker & breathing
- Marker: R marker
- Breathing: Suspend respiration.
Exposure by patient size
Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.
| Patient size | kV | mAs (chart) | mAs (DR default) |
|---|---|---|---|
| XS | 70 | 32 | 12.5 |
| S | 70 | 50 | 20 |
| M | 70 | 63 | 25 |
| L | 70 | 100 | 40 |
These start from a classic film-screen chart (70 kV, 100 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter. 80 kV with contrast: 50 mAs.
Image quality checklist
- The bladder is projected clear of the symphysis.
- The whole pelvic inlet is included.
- The anatomical side marker is visible, correct and clear of the anatomy.
- Collimation borders are visible on all sides, no wider than the area requested.
Faults & how to fix them
| Fault | Cause | Fix |
|---|---|---|
| Bladder overlaps the symphysis | No caudal angle | Angle 20° caudad. |
Common questions
- Where do you centre for a urinary bladder and inner pelvis AP X-ray?
- The centring point is 4 cm above the symphysis.
- What distance and beam angle does a urinary bladder and inner pelvis AP need?
- 100 cm from the tube to the receptor. The beam passes front to back (AP), tilted 20° towards the feet.
- What size receptor, and do you need a grid?
- Use a 24 × 30 cm receptor, landscape, in the table bucky. A grid is needed.
- What should the patient do with their breathing?
- Ask them to stop breathing for the moment of the exposure.
- What goes wrong most often on a urinary bladder and inner pelvis AP?
- Bladder overlaps the symphysis. Usual cause: no caudal angle. The fix: angle 20° caudad.
Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a urinary bladder and inner pelvis ap in a 3D room and score your positioning. See ScanFlo XR →
Written by the ScanFlo team from everyday radiographic practice.