Lateral knee X-ray positioning

One of the routine knee views. The patient lies on their side, the beam passes from side to side, straight on to the receptor, and the centring point is the medial femoral condyle.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Knee

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Lateral recumbent. On the side examined, the knee bent about 20–30°, a pad under the ankle; the other leg bent further forward.

Centring & collimation

Central ray

lateral

Centre: the medial femoral condyle.

Beam: lateral, perpendicular to the receptor.

Collimate: the joint with a quarter of femur and tibia

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmportrait
PlacementTabletopNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS531.60.8
S532.51.25
M533.21.6
L5352.5

These start from a classic film-screen chart (53 kV, 40 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The femoral condyles are superimposed.
  • The patellofemoral joint is open.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Condyles separateLeg rotated or ankle lowPad the ankle; square the knee.

Common questions

Where do you centre for a knee lateral X-ray?
The centring point is the medial femoral condyle.
What distance and beam angle does a knee lateral need?
100 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, portrait, on the tabletop. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a knee lateral?
Condyles separate. Usual cause: leg rotated or ankle low. The fix: pad the ankle; square the knee.