Knee tunnel view X-ray positioning (intercondylar)

An extra knee view, used when a routine one is not enough. The patient lies on their back, the beam passes front to back (AP), tilted 30° towards the feet, and the centring point is the knee joint.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Knee

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Supine with the knee flexed over a detector on a firm 12–15 cm support.

Centring & collimation

Central ray

AP

Centre: the knee joint.

Beam: AP, angled 30° caudad.

Collimate: the intercondylar fossa

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmlandscape
PlacementOn a raised supportNo grid
Angulation30°caudad

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS531.250.5
S5320.8
M532.51
L5341.6

These start from a classic film-screen chart (53 kV, 32 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The intercondylar fossa is shown open.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Fossa closedAngle not perpendicular to the tibiaAdjust the angle.

Common questions

Where do you centre for a knee intercondylar (tunnel) X-ray?
The centring point is the knee joint.
What distance and beam angle does a knee intercondylar (tunnel) need?
100 cm from the tube to the receptor. The beam passes front to back (AP), tilted 30° towards the feet.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, landscape, on a raised support. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a knee intercondylar (tunnel)?
Fossa closed. Usual cause: angle not perpendicular to the tibia. The fix: adjust the angle.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a knee intercondylar (tunnel) in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.