Ankle mortise view X-ray positioning (internal oblique)
One of the routine ankle views. The patient lies on their back, the beam passes front to back (AP), straight on to the receptor, and the centring point is midway between the malleoli.
By the ScanFlo Editorial Team · Updated
In this section · X-ray positioning
- X-ray positioning guides
- Skull positioning
- Face & sinuses positioning
- Soft tissue neck positioning
- C-spine positioning
- Chest positioning
- Shoulder positioning
- Humerus positioning
- Elbow positioning
- Forearm positioning
- Wrist positioning
- Hand positioning
- T-spine positioning
- Abdomen positioning
- L-spine positioning
- Sacrum & SI joints positioning
- Pelvis positioning
- Hip positioning
- Femur positioning
- Knee positioning
- Tibia & fibula positioning
- Ankle positioning
- Foot positioning
Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.
Patient position
Supine. Heel on the detector, the foot turned in 15°.
Centring & collimation
Central ray
APCentre: midway between the malleoli.
Beam: AP, perpendicular to the receptor.
Collimate: the distal third of tibia and fibula to the talus
Distance, receptor & grid
| Setting | Value | Notes |
|---|---|---|
| SID | 100 cm | |
| Receptor | 18 × 24 cm | portrait |
| Placement | Tabletop | No grid |
| Angulation | None | Perpendicular to the receptor |
Marker & breathing
- Marker: R or L marker on the side nearest the detector
- Breathing: No breathing instruction needed.
Exposure by patient size
Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.
| Patient size | kV | mAs (chart) | mAs (DR default) |
|---|---|---|---|
| XS | 53 | 0.8 | 0.5 |
| S | 53 | 1.25 | 0.5 |
| M | 53 | 1.6 | 0.63 |
| L | 53 | 2.5 | 1 |
These start from a classic film-screen chart (53 kV, 20 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.
Image quality checklist
- The whole ankle mortise is open.
- The anatomical side marker is visible, correct and clear of the anatomy.
- Collimation borders are visible on all sides, no wider than the area requested.
Faults & how to fix them
| Fault | Cause | Fix |
|---|---|---|
| Mortise closed laterally | Too little rotation | Turn the foot in 15°. |
Common questions
- Where do you centre for an ankle internal oblique (mortise) X-ray?
- The centring point is midway between the malleoli.
- What distance and beam angle does an ankle internal oblique (mortise) need?
- 100 cm from the tube to the receptor. The beam passes front to back (AP), straight on to the receptor.
- What size receptor, and do you need a grid?
- Use an 18 × 24 cm receptor, portrait, on the tabletop. No grid is needed.
- Where does the lead marker go?
- R or L marker on the side nearest the detector.
- What goes wrong most often on an ankle internal oblique (mortise)?
- Mortise closed laterally. Usual cause: too little rotation. The fix: turn the foot in 15°.
Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a ankle internal oblique (mortise) in a 3D room and score your positioning. See ScanFlo XR →
Written by the ScanFlo team from everyday radiographic practice.