Ankle mortise view X-ray positioning (internal oblique)

One of the routine ankle views. The patient lies on their back, the beam passes front to back (AP), straight on to the receptor, and the centring point is midway between the malleoli.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Ankle

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Heel on the detector, the foot turned in 15°.

Centring & collimation

Central ray

AP

Centre: midway between the malleoli.

Beam: AP, perpendicular to the receptor.

Collimate: the distal third of tibia and fibula to the talus

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmportrait
PlacementTabletopNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS530.80.5
S531.250.5
M531.60.63
L532.51

These start from a classic film-screen chart (53 kV, 20 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The whole ankle mortise is open.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Mortise closed laterallyToo little rotationTurn the foot in 15°.

Common questions

Where do you centre for an ankle internal oblique (mortise) X-ray?
The centring point is midway between the malleoli.
What distance and beam angle does an ankle internal oblique (mortise) need?
100 cm from the tube to the receptor. The beam passes front to back (AP), straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, portrait, on the tabletop. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on an ankle internal oblique (mortise)?
Mortise closed laterally. Usual cause: too little rotation. The fix: turn the foot in 15°.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a ankle internal oblique (mortise) in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.