Lateral ankle X-ray positioning

One of the routine ankle views. The patient lies on their back, the beam passes from side to side, straight on to the receptor, and the centring point is the medial malleolus.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Ankle

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Rotate the leg outwards so the lateral side of the foot rests on the detector.

Centring & collimation

Central ray

lateral

Centre: the medial malleolus.

Beam: lateral, perpendicular to the receptor.

Collimate: distal tibia to the base of the fifth metatarsal

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmportrait
PlacementTabletopNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS530.80.5
S531.250.63
M531.60.8
L532.51.25

These start from a classic film-screen chart (53 kV, 20 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The talar dome is superimposed.
  • Base of the fifth metatarsal included.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Double talar domeLeg rotatedAdjust to a true lateral.

Common questions

Where do you centre for an ankle lateral X-ray?
The centring point is the medial malleolus.
What distance and beam angle does an ankle lateral need?
100 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, portrait, on the tabletop. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on an ankle lateral?
Double talar dome. Usual cause: leg rotated. The fix: adjust to a true lateral.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a ankle lateral in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.