External oblique ankle X-ray positioning

An extra ankle view, used when a routine one is not enough. The patient lies on their back, the beam passes front to back (AP), straight on to the receptor, and the centring point is midway between the malleoli.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Ankle

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. On request: heel on the detector, foot rotated out to 30°.

Centring & collimation

Central ray

AP

Centre: midway between the malleoli.

Beam: AP, perpendicular to the receptor.

Collimate: the ankle

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmportrait
PlacementTabletopNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS530.80.5
S531.250.5
M531.60.63
L532.51

These start from a classic film-screen chart (53 kV, 20 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The medial malleolus is shown in profile.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Too little rotationNo angle guideUse a 30° pad.

Common questions

Where do you centre for an ankle external oblique X-ray?
The centring point is midway between the malleoli.
What distance and beam angle does an ankle external oblique need?
100 cm from the tube to the receptor. The beam passes front to back (AP), straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, portrait, on the tabletop. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on an ankle external oblique?
Too little rotation. Usual cause: no angle guide. The fix: use a 30° pad.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a ankle external oblique in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.