One of the routine wrist views. The patient sits, the beam passes back to front (PA), straight on to the receptor, and the centring point is midway between the styloid processes.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Wrist

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Sitting. Seat the patient at the holder, palm down, fingers slightly flexed, wrist centred.

Centring & collimation

Central ray

PA

Centre: midway between the styloid processes.

Beam: PA, perpendicular to the receptor.

Collimate: distal forearm to mid-metacarpals

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmlandscape
PlacementOn the holderNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS530.50.5
S530.50.5
M530.50.5
L530.80.5

These start from a classic film-screen chart (53 kV, 6.3 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Carpals, radius and ulna shown without rotation.
  • Distal radio-ulnar joint open.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
RotationPalm not flatFlatten the palm.

Common questions

Where do you centre for a wrist PA X-ray?
The centring point is midway between the styloid processes.
What distance and beam angle does a wrist PA need?
100 cm from the tube to the receptor. The beam passes back to front (PA), straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, landscape, on the holder. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a wrist PA?
Rotation. Usual cause: palm not flat. The fix: flatten the palm.