ScanFlo

Lateral wrist X-ray positioning

One of the routine wrist views. The patient sits, the beam passes from side to side, straight on to the receptor, and the centring point is the radial styloid.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Wrist

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Sitting. Standing or sitting; the ulnar border on the detector, thumb uppermost.

Centring & collimation

Central ray

lateral

Centre: the radial styloid.

Beam: lateral, perpendicular to the receptor.

Collimate: distal forearm to mid-metacarpals

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmlandscape
PlacementOn the holderNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS530.50.5
S530.630.5
M530.80.5
L531.250.5

These start from a classic film-screen chart (53 kV, 10 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Radius and ulna superimposed.
  • The lunate sits in the radius.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Ulna projects behindOver-rotatedAdjust to a true lateral.

Common questions

Where do you centre for a wrist lateral X-ray?
The centring point is the radial styloid.
What distance and beam angle does a wrist lateral need?
100 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, landscape, on the holder. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a wrist lateral?
Ulna projects behind. Usual cause: over-rotated. The fix: adjust to a true lateral.