Lateral tibia and fibula X-ray positioning (horizontal beam)

An extra tibia & fibula view, used when a routine one is not enough. The patient lies on their back, the beam passes from side to side, straight on to the receptor with the tube horizontal, and the centring point is the mid-shaft.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Tibia & fibula

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Do not move the injured leg; leave any splint on. Detector vertical beside the leg; the uninjured knee bent away.

Centring & collimation

Central ray

lateral

Centre: the mid-shaft.

Beam: lateral, perpendicular to the receptor, horizontal beam.

Collimate: knee to ankle

Distance, receptor & grid

SettingValueNotes
SID100 cmHorizontal beam
Receptor18 × 43 cmportrait
PlacementErect (wall) buckyNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS531.250.5
S5320.8
M532.51
L5341.6

These start from a classic film-screen chart (53 kV, 32 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The joint closest to the injured site is on the image.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Joint missedDetector shortShift to the joint nearest the injury.

Common questions

Where do you centre for a lower leg lateral (horizontal beam) X-ray?
The centring point is the mid-shaft.
What distance and beam angle does a lower leg lateral (horizontal beam) need?
100 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor with the tube horizontal.
What size receptor, and do you need a grid?
Use an 18 × 43 cm receptor, portrait, in the wall bucky. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a lower leg lateral (horizontal beam)?
Joint missed. Usual cause: detector short. The fix: shift to the joint nearest the injury.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a lower leg lateral (horizontal beam) in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.