Towne's view of the skull X-ray positioning (occiput)

An extra skull view, used when a routine one is not enough. The patient lies on their back, the beam passes front to back (AP), tilted 30° towards the feet, and the centring point is 7–8 cm above the nasion on the midline.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Skull

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Remove dentures, earrings, hair clips and anything in the hair. Supine, head flat on a thin pad, orbitomeatal line perpendicular to the table.

Centring & collimation

Central ray

AP

Centre: 7–8 cm above the nasion on the midline.

Beam: AP, angled 30° caudad.

Collimate: the occiput and the whole vault

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor24 × 30 cmportrait
PlacementTable buckyGrid in
Angulation30°caudad

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS70104
S70166.3
M70208
L703212.5

These start from a classic film-screen chart (70 kV, 40 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The dorsum sellae is projected within the foramen magnum.
  • Occipital bone and petrous ridges symmetric.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Dorsum sellae above the foramen magnumToo little caudal angle or chin upTuck the chin and check 30° caudad.

Common questions

Where do you centre for a skull (occiput) Towne's X-ray?
The centring point is 7–8 cm above the nasion on the midline.
What distance and beam angle does a skull (occiput) Towne's need?
100 cm from the tube to the receptor. The beam passes front to back (AP), tilted 30° towards the feet.
What size receptor, and do you need a grid?
Use a 24 × 30 cm receptor, portrait, in the table bucky. A grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a skull (occiput) Towne's?
Dorsum sellae above the foramen magnum. Usual cause: too little caudal angle or chin up. The fix: tuck the chin and check 30° caudad.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a skull (occiput) towne's in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.