ScanFlo

Towne's view of the mastoids X-ray positioning

An extra skull view, used when a routine one is not enough. The patient lies on their back, the beam passes front to back (AP), tilted 30° towards the feet, and the centring point is the midline between the external auditory meati.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Skull

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Remove dentures, earrings, hair clips and anything in the hair. Supine as for the occipital Towne's view, collimated to the petrous temporal bones.

Centring & collimation

Central ray

AP

Centre: the midline between the external auditory meati.

Beam: AP, angled 30° caudad.

Collimate: both petrous temporal bones

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmlandscape
PlacementTable buckyGrid in
Angulation30°caudad

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS70104
S70166.3
M70208
L703212.5

These start from a classic film-screen chart (70 kV, 40 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Both mastoids and petrous ridges symmetric.
  • Internal auditory meati visible.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Asymmetric petrous ridgesRotationCheck the mid-sagittal plane.

Common questions

Where do you centre for a mastoids Towne's X-ray?
The centring point is the midline between the external auditory meati.
What distance and beam angle does a mastoids Towne's need?
100 cm from the tube to the receptor. The beam passes front to back (AP), tilted 30° towards the feet.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, landscape, in the table bucky. A grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a mastoids Towne's?
Asymmetric petrous ridges. Usual cause: rotation. The fix: check the mid-sagittal plane.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a mastoids towne's in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.