AP shoulder X-ray positioning (trauma)

An extra shoulder view, used when a routine one is not enough. The patient sits, the beam passes front to back (AP), straight on to the receptor, and the centring point is the coracoid process.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Shoulder

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Sitting. Seat the patient against the bucky with the forearm across the body in its sling, pointing toward the tube.

Centring & collimation

Central ray

AP

Centre: the coracoid process.

Beam: AP, perpendicular to the receptor.

Collimate: glenohumeral joint and upper humerus

Distance, receptor & grid

SettingValueNotes
SID110 cm
Receptor24 × 30 cmlandscape
PlacementErect (wall) buckyGrid in
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: Suspend respiration.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS7020.8
S703.21.25
M7041.6
L706.32.5

These start from a classic film-screen chart (70 kV, 6.3 mAs at 140 cm) and are adjusted for a digital detector at 110 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Glenohumeral joint and upper humerus shown.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Arm movedSling removedLeave the arm in its sling.

Common questions

Where do you centre for a shoulder AP (after injury) X-ray?
The centring point is the coracoid process.
What distance and beam angle does a shoulder AP (after injury) need?
110 cm from the tube to the receptor. The beam passes front to back (AP), straight on to the receptor.
What size receptor, and do you need a grid?
Use a 24 × 30 cm receptor, landscape, in the wall bucky. A grid is needed.
What should the patient do with their breathing?
Ask them to stop breathing for the moment of the exposure.
What goes wrong most often on a shoulder AP (after injury)?
Arm moved. Usual cause: sling removed. The fix: leave the arm in its sling.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a shoulder ap (after injury) in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.