AP scapula X-ray positioning

One of the routine shoulder views. The patient lies on their back, the beam passes front to back (AP), straight on to the receptor, and the centring point is 5 cm below the coracoid.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Shoulder

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Supine with a pillow under the opposite shoulder; the arm abducted slightly to draw the scapula off the ribs.

Centring & collimation

Central ray

AP

Centre: 5 cm below the coracoid.

Beam: AP, perpendicular to the receptor.

Collimate: the whole scapula

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor24 × 30 cmportrait
PlacementTable buckyGrid in
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: Quiet breathing.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS703.21.25
S7052
M706.32.5
L70104
Child533.21.25

These start from a classic film-screen chart (70 kV, 12.5 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The scapula is clear of the ribs laterally.
  • Inferior angle included.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Inferior angle cut offCentring highCentre lower.

Common questions

Where do you centre for a scapula AP X-ray?
The centring point is 5 cm below the coracoid.
What distance and beam angle does a scapula AP need?
100 cm from the tube to the receptor. The beam passes front to back (AP), straight on to the receptor.
What size receptor, and do you need a grid?
Use a 24 × 30 cm receptor, portrait, in the table bucky. A grid is needed.
What should the patient do with their breathing?
They can keep breathing gently; no breath-hold is needed.
What goes wrong most often on a scapula AP?
Inferior angle cut off. Usual cause: centring high. The fix: centre lower.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a scapula ap in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.