PA oblique sternum X-ray positioning

An extra chest view, used when a routine one is not enough. The patient lies face down, turned at an angle, the beam passes back to front (PA), straight on to the receptor, and the centring point is the mid-sternum.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Chest

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Oblique, prone. Lay the patient prone, rotated 15–20° to project the sternum over the heart shadow.

Centring & collimation

Central ray

PA

Centre: the mid-sternum.

Beam: PA, perpendicular to the receptor.

Collimate: the sternum only

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor24 × 30 cmportrait
PlacementTable buckyGrid in
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: Quiet breathing.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS706.36.3
S701010
M7012.512.5
L702020

These start from a classic film-screen chart (70 kV, 25 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The sternum is projected over the heart, clear of the spine.
  • Lung markings are blurred by quiet breathing.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Sternum over the spineToo little rotationIncrease the obliquity.

Common questions

Where do you centre for a sternum oblique PA X-ray?
The centring point is the mid-sternum.
What distance and beam angle does a sternum oblique PA need?
100 cm from the tube to the receptor. The beam passes back to front (PA), straight on to the receptor.
What size receptor, and do you need a grid?
Use a 24 × 30 cm receptor, portrait, in the table bucky. A grid is needed.
What should the patient do with their breathing?
They can keep breathing gently; no breath-hold is needed.
What goes wrong most often on a sternum oblique PA?
Sternum over the spine. Usual cause: too little rotation. The fix: increase the obliquity.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a sternum oblique pa in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.