AP chest X-ray positioning (sitting)

An extra chest view, used when a routine one is not enough. The patient sits, the beam passes front to back (AP), straight on to the receptor, and the centring point is the sternum, midway between the jugular notch and xiphoid.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Chest

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Sitting. Sit the patient upright with the back against the bucky, shoulders down and rolled forward, chin raised. Use this only when the patient cannot stand.

Centring & collimation

Central ray

AP

Centre: the sternum, midway between the jugular notch and xiphoid.

Beam: AP, perpendicular to the receptor.

Collimate: apices to costophrenic angles, skin margin to skin margin

Distance, receptor & grid

SettingValueNotes
SID180 cm
Receptor35 × 43 cmportrait
PlacementErect (wall) buckyGrid in
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R marker, upper outer corner
  • Breathing: Breathe in and hold.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS1201.251
S12021.6
M1202.52
L12043.2
Child9021.6

These start from a classic film-screen chart (120 kV, 2.5 mAs at 140 cm) and are adjusted for a digital detector at 180 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The patient is upright: no fluid level is distorted by a slumped posture.
  • Both apices and costophrenic angles are included.
  • Clavicles are symmetric about the spine.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Heart looks enlargedAP projection magnifies the heartNote AP on the image; do not compare heart size with a PA.
Apices cut offPatient slumped down the buckySit the patient fully upright and recentre.

Common questions

Where do you centre for a chest AP sitting X-ray?
The centring point is the sternum, midway between the jugular notch and xiphoid.
What distance and beam angle does a chest AP sitting need?
180 cm from the tube to the receptor. The beam passes front to back (AP), straight on to the receptor.
What size receptor, and do you need a grid?
Use a 35 × 43 cm receptor, portrait, in the wall bucky. A grid is needed.
What should the patient do with their breathing?
Ask them to breathe in and hold it while you expose.
What goes wrong most often on a chest AP sitting?
Heart looks enlarged. Usual cause: aP projection magnifies the heart. The fix: note AP on the image; do not compare heart size with a PA.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a chest ap sitting in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.