AP coccyx X-ray positioning

An extra sacrum & SI joints view, used when a routine one is not enough. The patient lies on their back, the beam passes front to back (AP), tilted 10° towards the feet, and the centring point is 5 cm above the symphysis.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Sacrum & SI joints

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Supine, bladder emptied, legs straight.

Centring & collimation

Central ray

AP

Centre: 5 cm above the symphysis.

Beam: AP, angled 10° caudad.

Collimate: the coccyx

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmportrait
PlacementTable buckyGrid in
Angulation10°caudad

Marker & breathing

  • Marker: anatomical side marker on the side being examined, clear of anatomy
  • Breathing: Suspend respiration.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS70104
S70166.3
M70208
L703212.5

These start from a classic film-screen chart (70 kV, 40 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The coccyx is clear of the symphysis.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Coccyx behind the symphysisNo caudal angleAngle 10° caudad.

Common questions

Where do you centre for a coccyx AP X-ray?
The centring point is 5 cm above the symphysis.
What distance and beam angle does a coccyx AP need?
100 cm from the tube to the receptor. The beam passes front to back (AP), tilted 10° towards the feet.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, portrait, in the table bucky. A grid is needed.
What should the patient do with their breathing?
Ask them to stop breathing for the moment of the exposure.
What goes wrong most often on a coccyx AP?
Coccyx behind the symphysis. Usual cause: no caudal angle. The fix: angle 10° caudad.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a coccyx ap in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.