PA finger X-ray positioning

One of the routine hand views. The patient sits, the beam passes back to front (PA), straight on to the receptor, and the centring point is the proximal interphalangeal joint.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Hand

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Sitting. Palm down, the finger extended and slightly separated.

Centring & collimation

Central ray

PA

Centre: the proximal interphalangeal joint.

Beam: PA, perpendicular to the receptor.

Collimate: the finger and its metacarpal head

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmportrait
PlacementOn the holderNo grid
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS460.50.5
S460.50.5
M460.50.5
L460.80.5

These start from a classic film-screen chart (46 kV, 8 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Joints open, no rotation.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
RotationFinger turnedLay the finger flat.

Common questions

Where do you centre for a finger PA X-ray?
The centring point is the proximal interphalangeal joint.
What distance and beam angle does a finger PA need?
100 cm from the tube to the receptor. The beam passes back to front (PA), straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, portrait, on the holder. No grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a finger PA?
Rotation. Usual cause: finger turned. The fix: lay the finger flat.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a finger pa in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.