Lateral femur X-ray positioning (trauma)

An extra femur view, used when a routine one is not enough. The patient lies on their back, the beam passes from side to side, straight on to the receptor with the tube horizontal, and the centring point is the mid-shaft.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Femur

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Supine. Do not move the injured leg. Detector vertical against it on a foam pad; raise and support the other leg.

Centring & collimation

Central ray

lateral

Centre: the mid-shaft.

Beam: lateral, perpendicular to the receptor, horizontal beam.

Collimate: the knee and the shaft

Distance, receptor & grid

SettingValueNotes
SID100 cmHorizontal beam
Receptor18 × 43 cmportrait
PlacementErect (wall) buckyGrid in
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: Suspend respiration.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS706.32.5
S70104
M7012.55
L70208

These start from a classic film-screen chart (70 kV, 25 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • The knee joint is shown.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Knee missedCentring highMake sure the knee is on the film.

Common questions

Where do you centre for a femur lateral (after injury) X-ray?
The centring point is the mid-shaft.
What distance and beam angle does a femur lateral (after injury) need?
100 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor with the tube horizontal.
What size receptor, and do you need a grid?
Use an 18 × 43 cm receptor, portrait, in the wall bucky. A grid is needed.
What should the patient do with their breathing?
Ask them to stop breathing for the moment of the exposure.
What goes wrong most often on a femur lateral (after injury)?
Knee missed. Usual cause: centring high. The fix: make sure the knee is on the film.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a femur lateral (after injury) in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.