Occipitomental (Waters) view X-ray positioning (sinuses and face)

One of the routine face & sinuses views. The patient sits, the beam passes back to front (PA), straight on to the receptor, and the centring point is the acanthion.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/Face & sinuses

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Sitting. Remove dentures, earrings, hair clips and anything in the hair. Seat the patient facing the bucky, mouth wide open, chin on the bucky, head tilted back so the orbitomeatal line is at 45°.

Centring & collimation

Central ray

PA

Centre: the acanthion.

Beam: PA, perpendicular to the receptor.

Collimate: frontal to maxillary sinuses

Distance, receptor & grid

SettingValueNotes
SID100 cm
Receptor18 × 24 cmportrait
PlacementErect (wall) buckyGrid in
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: No breathing instruction needed.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS8083.2
S8012.55
M80166.3
L802510

These start from a classic film-screen chart (80 kV, 25 mAs at 140 cm) and are adjusted for a digital detector at 100 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • Petrous ridges just below the floors of the maxillary sinuses.
  • Sphenoid sinus seen through the open mouth.
  • Fluid levels are horizontal (erect).
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
Petrous ridges over the maxillary sinusesHead not tilted back enoughRaise the chin to 45°.

Common questions

Where do you centre for a sinuses and face semiaxial (occipitomental) X-ray?
The centring point is the acanthion.
What distance and beam angle does a sinuses and face semiaxial (occipitomental) need?
100 cm from the tube to the receptor. The beam passes back to front (PA), straight on to the receptor.
What size receptor, and do you need a grid?
Use an 18 × 24 cm receptor, portrait, in the wall bucky. A grid is needed.
Where does the lead marker go?
R or L marker on the side nearest the detector.
What goes wrong most often on a sinuses and face semiaxial (occipitomental)?
Petrous ridges over the maxillary sinuses. Usual cause: head not tilted back enough. The fix: raise the chin to 45°.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a sinuses and face semiaxial (occipitomental) in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.