Lateral cervical spine X-ray positioning (flexion)

An extra C-spine view, used when a routine one is not enough. The patient stands side-on, the beam passes from side to side, straight on to the receptor, and the centring point is C4.

By the ScanFlo Editorial Team · Updated

In this section · X-ray positioning

X-ray positioning/C-spine

Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.

Patient position

Lateral, erect. Only on request and not after injury: lateral as standard, chin brought down onto the chest.

Centring & collimation

Central ray

lateral

Centre: C4.

Beam: lateral, perpendicular to the receptor.

Collimate: base of skull to T1

Distance, receptor & grid

SettingValueNotes
SID150 cm
Receptor24 × 30 cmportrait
PlacementErect (wall) buckyGrid in
AngulationNonePerpendicular to the receptor

Marker & breathing

  • Marker: R or L marker on the side nearest the detector
  • Breathing: Suspend respiration.

Exposure by patient size

Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.

Patient sizekVmAs (chart)mAs (DR default)
XS7083.2
S7012.55
M70166.3
L702510

These start from a classic film-screen chart (70 kV, 12.5 mAs at 140 cm) and are adjusted for a digital detector at 150 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.

Image quality checklist

  • All cervical bodies shown in flexion.
  • Labelled flexion.
  • The anatomical side marker is visible, correct and clear of the anatomy.
  • Collimation borders are visible on all sides, no wider than the area requested.

Faults & how to fix them

FaultCauseFix
C7 hiddenShoulders highDepress the shoulders.

Common questions

Where do you centre for a cervical spine lateral in flexion X-ray?
The centring point is C4.
What distance and beam angle does a cervical spine lateral in flexion need?
150 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor.
What size receptor, and do you need a grid?
Use a 24 × 30 cm receptor, portrait, in the wall bucky. A grid is needed.
What should the patient do with their breathing?
Ask them to stop breathing for the moment of the exposure.
What goes wrong most often on a cervical spine lateral in flexion?
C7 hidden. Usual cause: shoulders high. The fix: depress the shoulders.

Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a cervical spine lateral in flexion in a 3D room and score your positioning. See ScanFlo XR →

Written by the ScanFlo team from everyday radiographic practice.