Lateral pregnancy abdomen X-ray positioning (erect)
An extra abdomen view, used when a routine one is not enough. The patient stands side-on, the beam passes from side to side, straight on to the receptor, and the centring point is the pelvis, including the symphysis.
By the ScanFlo Editorial Team · Updated
In this section · X-ray positioning
- X-ray positioning guides
- Skull positioning
- Face & sinuses positioning
- Soft tissue neck positioning
- C-spine positioning
- Chest positioning
- Shoulder positioning
- Humerus positioning
- Elbow positioning
- Forearm positioning
- Wrist positioning
- Hand positioning
- T-spine positioning
- Abdomen positioning
- L-spine positioning
- Sacrum & SI joints positioning
- Pelvis positioning
- Hip positioning
- Femur positioning
- Knee positioning
- Tibia & fibula positioning
- Ankle positioning
- Foot positioning
Still being checked: a radiographer is reviewing this guide. If your department does it differently, follow your department.
Patient position
Lateral, erect. Only after week 37 and only if ultrasound is unavailable. Bladder emptied; patient stands left side to the bucky with a measuring rod between the legs.
Centring & collimation
Central ray
lateralCentre: the pelvis, including the symphysis.
Beam: lateral, perpendicular to the receptor.
Collimate: pelvis and lower abdomen
Distance, receptor & grid
| Setting | Value | Notes |
|---|---|---|
| SID | 110 cm | |
| Receptor | 35 × 43 cm | portrait |
| Placement | Erect (wall) bucky | Grid in |
| Angulation | None | Perpendicular to the receptor |
Marker & breathing
- Marker: L marker
- Breathing: Suspend respiration.
Exposure by patient size
Treat these as a starting point. Your detector, your generator and the patient in front of you will all move the numbers.
| Patient size | kV | mAs (chart) | mAs (DR default) |
|---|---|---|---|
| XS | 90 | 25 | 10 |
| S | 90 | 40 | 16 |
| M | 90 | 50 | 20 |
| L | 90 | 80 | 32 |
These start from a classic film-screen chart (90 kV, 63 mAs at 140 cm) and are adjusted for a digital detector at 110 cm. The DR column is what ScanFlo's console suggests; each step up or down changes the mAs by about a quarter.
Image quality checklist
- The symphysis and the measuring rod are shown.
- Justification recorded: no ultrasound available.
- The anatomical side marker is visible, correct and clear of the anatomy.
- Collimation borders are visible on all sides, no wider than the area requested.
Faults & how to fix them
| Fault | Cause | Fix |
|---|---|---|
| Unjustified exposure | Ultrasound available | Do not perform; use ultrasound. |
Common questions
- Where do you centre for a pregnancy lateral erect X-ray?
- The centring point is the pelvis, including the symphysis.
- What distance and beam angle does a pregnancy lateral erect need?
- 110 cm from the tube to the receptor. The beam passes from side to side, straight on to the receptor.
- What size receptor, and do you need a grid?
- Use a 35 × 43 cm receptor, portrait, in the wall bucky. A grid is needed.
- What should the patient do with their breathing?
- Ask them to stop breathing for the moment of the exposure.
- What goes wrong most often on a pregnancy lateral erect?
- Unjustified exposure. Usual cause: ultrasound available. The fix: do not perform; use ultrasound.
Try it in ScanFlo: the ScanFlo XR X-ray Room, coming soon, lets you set up a pregnancy lateral erect in a 3D room and score your positioning. See ScanFlo XR →
Written by the ScanFlo team from everyday radiographic practice.